SG700 Stack Gas Analyzer

The SG700 Stack Gas Analyzer consists of a non-dispersive infrared gas analyzer using a highly sensitive and stable double beam system and a zirconia or paramagnetic oxygen analyzer. It can simultaneously measure up to 5 components: NOx, SO2, CO2, CO, and O2.

The use of interference compensation in the analyzing section virtually eliminates the effects of interference from other components. The SG700 features many functions including autocalibration, conversion to O2-based values, averaging, and high/low alarms. It is used for measuring and monitoring emissions from various types of boilers and waste incinerators.



Features

* Simultaneously measure up to five gas components:
* Minimize interaction between gases
* Reliable zirconia oxygen analyzer
* A wealth of functions, including self-diagnostic functions which enhance reliability
* Wide dynamic range



Specification

Measurement Up to 4 components (NOx, SO2, CO2, CO) and O2
Measurering method
NOx, SO2, CO2, CO: Non-dispersive Infrared Absorption (NDIR)
O2: Zirconia or paramagnetic Cell method
Measuring ranges
NOx: :0-50 ppm to 0-5000 ppm
SO2: 0-50 ppm to 0-1000 ppm
CO2: 0-1 vol% to 0-20 vol%
CO: 0-50 ppm to 0-5000 ppm
O2: 0-10/0-25 vol%
Display LCD with backlight
Output signal 5 points, 4-20mA DC or 0 to 1 V DC
Contact output Each component range identification, analyzer failure, calibration failure, calibration status, maintenance status, CO peak count alarm, each instantaneous value alarm, analyzer power shutdown
Contact input Auto calibration start, average reset, Range switching output hold, remote pump off
Construction Outdoor/indoor stand-alone system, non-explosionproof, rainproof
Ambient temperature -5 to 40 degC
Ambient humidity 90%RH or less
Power supply 100/110/115/200/230 V AC ±10%, 50/60 Hz ±0.5 Hz
Power consumption Approx. 600 to 1000 VA
Weight Approx. 350 kg (without standard gases)
Sample gas conditions
Temperature: 1400 degC or less
Dust: 500 mg/Nm3 or less
Pressure: -5 to 5 kPa
Flow rate: Approx. 2 l/min
Measurement Up to 4 components (NOx, SO2, CO2, CO) and O2
Measurering method
NOx, SO2, CO2, CO: Non-dispersive Infrared Absorption (NDIR)
O2: Zirconia or paramagnetic Cell method
Measuring ranges
NOx: :0-50 ppm to 0-5000 ppm
SO2: 0-50 ppm to 0-1000 ppm
CO2: 0-1 vol% to 0-20 vol%
CO: 0-50 ppm to 0-5000 ppm
O2: 0-10/0-25 vol%
Display LCD with backlight
Output signal 5 points, 4-20mA DC or 0 to 1 V DC
Contact output Each component range identification, analyzer failure, calibration failure, calibration status, maintenance status, CO peak count alarm, each instantaneous value alarm, analyzer power shutdown
Contact input Auto calibration start, average reset, Range switching output hold, remote pump off
Construction Outdoor/indoor stand-alone system, non-explosionproof, rainproof
Ambient temperature -5 to 40 degC
Ambient humidity 90%RH or less
Power supply 100/110/115/200/230 V AC ±10%, 50/60 Hz ±0.5 Hz
Power consumption Approx. 600 to 1000 VA
Weight Approx. 350 kg (without standard gases)
Sample gas conditions
Temperature: 1400 degC or less
Dust: 500 mg/Nm3 or less
Pressure: -5 to 5 kPa
Flow rate: Approx. 2 l/min