The SG700 Stack Gas Analyzer consists of a non-dispersive infrared gas analyzer using a highly sensitive and stable double beam system and a zirconia or paramagnetic oxygen analyzer. It can simultaneously measure up to 5 components: NOx, SO
2, CO
2, CO, and O
2.
The use of interference compensation in the analyzing section virtually eliminates the effects of interference from other components. The SG700 features many functions including autocalibration, conversion to O
2-based values, averaging, and high/low alarms. It is used for measuring and monitoring emissions from various types of boilers and waste incinerators.
Features
* | Simultaneously measure up to five gas components: |
* | Minimize interaction between gases |
* | Reliable zirconia oxygen analyzer |
* | A wealth of functions, including self-diagnostic functions which enhance reliability |
* | Wide dynamic range |
Specification
Measurement | Up to 4 components (NOx, SO2, CO2, CO) and O2 |
Measurering method | NOx, SO2, CO2, CO: Non-dispersive Infrared Absorption (NDIR) | O2: Zirconia or paramagnetic Cell method | |
Measuring ranges | NOx: | :0-50 ppm to 0-5000 ppm | SO2: | 0-50 ppm to 0-1000 ppm | CO2: | 0-1 vol% to 0-20 vol% | CO: | 0-50 ppm to 0-5000 ppm | O2: | 0-10/0-25 vol% | |
Display | LCD with backlight |
Output signal | 5 points, 4-20mA DC or 0 to 1 V DC |
Contact output | Each component range identification, analyzer failure, calibration failure, calibration status, maintenance status, CO peak count alarm, each instantaneous value alarm, analyzer power shutdown |
Contact input | Auto calibration start, average reset, Range switching output hold, remote pump off |
Construction | Outdoor/indoor stand-alone system, non-explosionproof, rainproof |
Ambient temperature | -5 to 40 degC |
Ambient humidity | 90%RH or less |
Power supply | 100/110/115/200/230 V AC ±10%, 50/60 Hz ±0.5 Hz |
Power consumption | Approx. 600 to 1000 VA |
Weight | Approx. 350 kg (without standard gases) |
Sample gas conditions | Temperature: 1400 degC or less | Dust: 500 mg/Nm3 or less | Pressure: -5 to 5 kPa | Flow rate: Approx. 2 l/min | |
Measurement | Up to 4 components (NOx, SO2, CO2, CO) and O2 |
Measurering method | NOx, SO2, CO2, CO: Non-dispersive Infrared Absorption (NDIR) | O2: Zirconia or paramagnetic Cell method | |
Measuring ranges | NOx: | :0-50 ppm to 0-5000 ppm | SO2: | 0-50 ppm to 0-1000 ppm | CO2: | 0-1 vol% to 0-20 vol% | CO: | 0-50 ppm to 0-5000 ppm | O2: | 0-10/0-25 vol% | |
Display | LCD with backlight |
Output signal | 5 points, 4-20mA DC or 0 to 1 V DC |
Contact output | Each component range identification, analyzer failure, calibration failure, calibration status, maintenance status, CO peak count alarm, each instantaneous value alarm, analyzer power shutdown |
Contact input | Auto calibration start, average reset, Range switching output hold, remote pump off |
Construction | Outdoor/indoor stand-alone system, non-explosionproof, rainproof |
Ambient temperature | -5 to 40 degC |
Ambient humidity | 90%RH or less |
Power supply | 100/110/115/200/230 V AC ±10%, 50/60 Hz ±0.5 Hz |
Power consumption | Approx. 600 to 1000 VA |
Weight | Approx. 350 kg (without standard gases) |
Sample gas conditions | Temperature: 1400 degC or less | Dust: 500 mg/Nm3 or less | Pressure: -5 to 5 kPa | Flow rate: Approx. 2 l/min | |