IR400 Infrared Gas Analyzer

The IR200 infrared gas analyzer is capable of measuring the concentrations of CO2, CO, CH4, SO2, NO and O2 components in sample gas. CO2, CO, CH4, SO2 and NO are measured by the non-dispersion infrared method (NDIR), while O2 is measured by the paramagnetic or zirconia method. A maximum of 4 components including O2 (max. 3 components except for O2 measurement) are simultaneously measurable.

A high-sensitivity mass flow sensor is adopted in the detection block for the infrared method. Due to use of a single beam system for measurement, maintenance is easy and excellent stability is ensured for a long period of time. In addition, a microprocessor is built in and a large-size liquid crystal display is provided for easier operation, higher accuracy and more functions.

This analyzer is thus optimum for combustion control of various industrial furnaces, botanical study and global atmospheric research.



Features

* Simultaneous measurement of up to 4 components including O2 and 3 components selected from among CO2, CO, CH4, SO2, and NO.
* A unique optics system minimizes drift particularly due to contamination of measurement cell, ensuring excellent long-term stability.
* The dual cell type of transmission detector minimizes interference from other gas components.
* Single beam system allows for simple measurement unit construction and requires no adjustment of optical balance, resulting in low maintenance.
* Large LCD provides easy interactive operation.
* Highly precise zero/span calibration and automatic calibration are available. Self-diagnostic function detects abnormality and displays an error massage.



Specifications

Measurement principle CO2, CO, CH4, SO2, NO: Non-dispersive infrared method, Single light source-single beam
O2: Paramagnetic type (built-in), or zirconia type (external)
Sample gas conditions
Flow rate: 1 l/min±0.5 l/min
Temperature: 0 to 50 degC
Pressure: 10 kPa or less (gas outlet side should be open to the atmospheric)
Dust: 100μg/Nm3 or less in particle size of 0.3μm or less
Mist: Unallowable
Moisture: Below the level where saturation occurs at room temperature (condensation unallowable).
Corrosive component: HCl 1 ppm or less
Measuring gas components and measuring range

Minimum range Maximum range
CO2 0-500ppm 0-100vol%
CO 0-200ppm 0-100vol%
CH4 0-1000ppm 0-100vol%
SO2 0-500ppm 0-5000ppm
NO 0-500ppm 0-5000ppm
O2 (built in) 0-5vol% 0-100vol%
O2 (External Zirconia) 0-5vol% 0-25vol%
Analog output signals 4 to 20mA DC or 0 to 1V DC, nonisolated output
Analog input signal For signal input from externally installed O2 sensor. Nonisolated input.
Contact input Non-voltage contact (ON/0V, OFF/5V DC, 5mA flowing at ON)
Ambient temperature -5 to 45 degC
Ambient humidity 90% RH max., noncondensing
Dimensions (HxWxD) 19-inch rack mounting type; 177x483x493mm
Weight Approx. 10kg
Power supply 85 to 264 V AC, 50/60 Hz (3-pin inlet terminal used)